Installation and Operating Instructions OPTIMASS

47

Output Option 3

In addition to the 1

st

4...20mA, a control or binary input can be wired as shown.









0 V DC


Output Option 4

In addition to the 1

st

4...20 mA, a status or alarm output can be wired as shown.









Status

0 V DC


Output Option 5

The Profibus communication output available on this converter can be wired as shown in addition to a
4...20mA output. Refer to communication handbook for connection details










Profibus PA

5 … 30 V DC

Contact

5 … 30 V DC