Installation and Operating Instructions OPTIMASS

79

To set a time constant: e.g. 0.5 sec

Display

Key

Line 1

Line 2

From 3.1.(2) Time.

Const.

(0)0.3 Time.

Const.

(0).2 Time.

Const.

0.(2) Time.

Const.

Repeat

key till 5 is reached, then

to accept the value.

The standard range for the time constant is 0.2...20 seconds.

The filtering only applies to mass and volume flow readings and any outputs programmed to these values.

The mass totaliser, volume totaliser, density and temperature measurement outputs are independent of

the programmed time constant.

Flow Mode Fct. 3.1.3

This setting allows the user to select the option of measuring flow in one direction only or in both

directions.

To select the option required:

Display

Key

Line 1

Line 2

From 3.1.(3) Flow

mode

(Flow +/-)

default setting

The ↑ key allows the selection of one of the following:

Flow>0

Ignores negative flows

Flow<0

Ignores positive flows

Flow +/- Allows positive and negative flows

When the correct option is in the display the ↵ key may be pressed to accept the option.

Note:

The totaliser will increment AND decrement if flow ± is selected according to the flow direction. Status

options are available to indicate whether a reverse or forward flow is registered.