Installation and Operating Instructions OPTIMASS

48

2.6 Compact to Remote / Remote to Compact conversion instructions


Conversion of the mounting of the signal converter from compact to remote, or vice-versa, is possible
under certain circumstances, using a kit of parts.


A change from compact to remote or vice versa in the field is possible in safe areas, for hazardous areas
only in the workshop.


Contact your local KROHNE Company and provide the serial number of the meter for further details.

2.7 Technical Data

2.7.1 MFC050

Power Supply

Operating Voltage :

115V AC (+10% / -15%)
230V AC (+10% / -15%)
24V DC (+/- 30%)

Power consumption :

AC – 18VA
DC – 10W


Inputs and Outputs

Current (mA) output

Function :

Active (converter powered)

Level :

0/4 … 20mA

Maximum Load :

500

Ω


Pulse Output

Function :

Passive (externally powered) open collector transistor switch

Maximum Frequency :

Frequency – 1300Hz, Pulse – 10 KHz

Pulse width :

0.05 … 500 mS (settable)

External power supply :

<24V DC

Maximum circuit current : < 150mA

Control Input

Function :

Passive (externally powered)

Input signal state high :

4 … 24V DC

Input signal state low :

< 2V DC or open circuit


Status Output

Function :

Active (converter powered) or
Passive (externally powered) depending on output options fitted

Active configuration

Output signal state high :

24V DC

Output maximum current :

20mA

Passive configuration

External circuit voltage :

< 24V DC

Maximum circuit current :

20mA