Installation and Operating Instructions OPTIMASS

75

Testing the pulse output Fct. 2.5

To test the pulse output, connect an external counter to the output terminals. When testing the pulse

output, the operator has the choice of the following pulse widths: 0.4 ms, 1.0 ms, 10.0 ms, 100.0 ms and

500 ms.

The operator should choose the pulse width that best matches the performance of his external pulse

counter.

Connect a pulse counter to the impulse output and proceed as follows:

Display

Key

Line 1

Line 2

→↑

Fct. (2)

TEST

Fct. 2.(1)

DISPLAY

x 4

Fct. 2.(5)

Pulse out

SURE

(NO)

SURE

(YES)

Select pulse width with ↑ key

Start test pulse output

The meter now issues a stream of pulses with the set width. The running total of pulses sent is shown on

the display. The test stops when either 100,000 pulses have been sent or the operator presses the ↵ key.

A connected counter will now count. Press ↵ to stop the counting. The meter display and counter should

have matching totals.

If the counter reads a smaller number than the actual number of pulses sent, or the frequency meter under

reads, then this indicates that a weak signal is reaching the frequency meter/pulse counter. In this case try

the following suggestions:

Decrease the external pull-up resistor

Decrease/remove the filter capacitor.

Decrease the cable length between the converter and the counter.

Add external buffers to boost the signal.

If the pulse counter reads a larger number than the converter, or if the frequency meter reading is high or

unstable, then this indicates the presence of external interference. Try one or more of the following:

Add/increase the size of the filter capacitor. (10...100nF)

Use high quality screened cable.

Keep cable lengths as short as possible, avoiding high power equipment/switchgear and any cabling

connected to them.

Use external buffers.


Testing the Frequency Output Fct. 2.6

This function allows the frequency output to be tested. The Frequency output has an open collector

transistor drive which requires a pull-up resistor to an external DC power supply. To test the frequency,

connect a frequency counter to the output terminals and proceed as follows:

Display

Key

Line 1

Line 2

→↑

Fct. (2)

TEST

Fct. 2.(1)

DISPLAY

x 5

Fct. 2.(6)

Freq. out

SURE

(NO)

SURE

(YES)

(Level

Low)

0V on the output

(Level

High)

+24V on the output

1

Hz

10

Hz

100

Hz

1000

Hz

A frequency counter connected to the output will show these frequencies in steps.

Return to Fct. 2.(6)