Installation and Operating Instructions OPTIMASS

7

Examples

Vertical mounting

Horizontal mounting

Upward sloping installation

Avoid mounting the meter with long vertical

drops after the meter. This could cause

siphoning and cause measurement errors.

Avoid mounting the meter at the highest point in

the pipeline. Air or gas can accumulate here and

cause faulty measurements

1

2

Flow

2

1

Flow

1

Valve for zeroing flow meter

2

The second valve is recommended if the pump is switched off to prevent reverse flow

To enable a good zero to be done, it is recommended that a shut-off valve be installed downstream

of the flow meter.