1.4.4 Secondary Pressure containment
The OPTIMASS 8000/9000 series sensors do not have certified secondary containment.

If the user suspects that the primary tube has failed, the unit must be depressurised and removed from

service as soon as possible.


1.4.5 Pressure de-rating
Meter data plates are stamped with maximum pressure rating (at max. operating temperature) of

connection, primary tube or secondary pressure containment (whichever is the lower). Higher pressures

may be possible at lower temperatures.

Measuring Tubes:

Process Temperature

Process Temperature Process

Temperature

Maximum 150 ºC / 300 ºF Maximum 230 ºC / 440 ºF Maximum 350 ºC / 660 ºF

(9000 series only)

Meter

size

barg

psig

barg

psig

barg

psig

15

210

3045

185

2680

160

2320

25

165

2390

145

2100

125

1810

40

140

2030

120

1740

105

1520

80

125

1810

110

1595

95

1375

100

85

1230

75

1085

65

940

Flanges:

DIN flanges to EN1092-1. Note also pressure / temperature limits for measuring tubes above.