Page 70

background image

Installation and Operating Instructions OPTIMASS

69

Under certain conditions, it may not be possible to adjust the zero point, for instance when:

The medium is in motion, because the shut-off valves etc. are not functioning properly.

There are still gaseous inclusions in the primary head because it was flushed insufficiently.

In such cases the zero point adjustment will not be accepted. If the zero adjustment was started by the

binary input, the converter will show the message :

ZERO.ERROR

The converter also reports the ZERO.ERROR in the status list.

Under certain circumstances, when the media consists of unevenly mixed components, it might be

difficult to adjust the zero point. In such a case, the zero point adjustment procedure must be carried

out under special conditions:

Media which tend to vaporise or degas should be kept under higher pressure.

Two-phase media consisting of a separable solid component (slurry): In such a case it might be

advisable to fill the primary head with the carrier medium only.

Other two-phase media.

If it is not possible to separate the solid or gaseous components, the operator can fill the measuring

system with a substitute liquid (e.g. with water).


Instrument State Fct. 1.2.

The instrument may be switched to a 'STANDBY' state. Once in this state, all outputs go to their off

state and the mass totaliser is frozen. The main display will have the STANDBY indicator set and will

display either the frozen totaliser or just STANDBY.

Begin from measuring mode

Display

Key

Line 1

Line 2

STANDBY

3.456

kg (Frozen Totalizer)

STANDBY

Whilst in this state the measuring tube still vibrates and the measurements can come back on line in

an instant.

There is an additional standby state, 'STOP', in this case the drive to the primary head is disabled and

vibrations cease. However, when leaving STOP the converter has to return to STARTUP before

measurements can resume.

The instrument can be switched to STANDBY or STOP either by the keys on the display or by the

control input signal (see section 5.4). STOP can only be set by the keys.

To set STANDBY or STOP:

Begin from measuring mode

Display

Key

Line 1

Line 2

x 2

Fct. 1.(1)

ZERO CAL.

Fct. 1.(2.).

INST. STATE

MEASURE

(STANDBY)

(STOP)

Use the ↑ key to select the desired mode.

Fct. 1.(2)

STANDBY

If STANDBY or STOP was selected the instrument goes immediately into that state.

To return to measurement, go back to Fct. 1.2 and select MEASURE.