Installation and Operating Instructions OPTIMASS

61

Fct.
No

Text Description

and

settings

3.2.11 CONC.BY.VOL. Monitor concentration by volume

* Select OFF (no concentration by volume on display or
* PERCENT V

3.2.12 VELOCITY

Monitor flow velocity
* Select OFF (no flow velocity on display or)
* m/sec
* ft/sec

3.2.13 LANGUAGE

Language for display text
* ENGLISH
* FRANCAIS
* ESPANOL
* DEUTSCH

3.3

CONC. MEAS.

Submenu 3.3 Concentration measurement

3.3.1

CONC. MODE

* NOT FITTED (no concentration measurement available) or select option (only if
concentration was ordered):
* NONE
* BRIX
* GEN. CONC.
* BAUME 144.3
* BAUME 145.0
* NAOH
* PLATO

3.3.2

ENABLE.CONC Enter the access code for concentration measurement

If access is enabled:

3.3.2 OFFSET

Offset for concentration measurement
Input manual concentration offset
* Direct input of a concentration offset.

3.3.3 CONC

TYPE

3.3.4 CONC

CF1

3.3.12 CONC

CF12

Refer to separate concentration manual

3.4 DENSITY

Submenu

3.4

Density

3.4.1 DENS.

MODE Option density mode

Press key

↵ , select with key → and ↑ unit and value, quit with key ↵ back to function

3.1.5.
FIXED (Norm density)
REFERRED (temperature-referred density)
ACTUAL (operating density)

3.4.2 FIXED

Input of the fixed density for the option „FIXED“ only

3.4.2 REF

TEMP

Input of the reference temperature for the option „REFERRED“ only

3.4.3 SLOPE

Input of the slope for the option „REFERRED“ only

3.5 PASSWORDS

Submenu

3.5

Passwords

3.5.1 SUPERVISOR Supervisor code for accessing menus required?

Use the key

↑ to select, then press ↵.

* ENABLE PW
* CHANGE PW (set a 9-keystroke code)
* EXIT
Default setting:

→ → → ↵ ↵ ↵ ↑ ↑ ↑

3.5.2 CUSTODY

Custody transfer code required?

3.5.3 TOTAL.RESET Totaliser reset enabled?

Use the key

↑ to select, then press ↵.

* ALLOW.RESET (reset unblocked)
* COMM RESET (reset via communication options is enabled)
* NO RESET (reset blocked)

3.6 SETTINGS

Submenu

3.6

Settings

3.6.1 TAG

ID.

Tag name setting

(measuring point number)

Required only for flow meters using the MIC 500 Hand Held Communicator (HHC),
connected to current output)..
Factory setting: ”MFC 050 (or MFC 51)”
Characters assignable to each place:
A...Z / 0...9 / + / - / * / = / // ( > = blank character)