Installation and Operating Instructions OPTIMASS

63

Fct.
No

Text Description

and

settings

4.5

PULSE OUT.

Submenu 4.5 Pulse/frequency output

Function pulse/frequency output P
* OFF (Output = 0V DC)
* MASS FLOW (Frequency output 0 to MAX Freq. Hz = Mass Flow in range:
MIN. FLOW to MAX FLOW as set in Fct. 4.5.2 and 4.5.3)
* DENSITY (Frequency output 0 to MAX Freq. Hz = Density in range:
MIN.DENSITY to MAX.DENSITY as set in Fct. 4.5.2 and 4.5.3)
* MASS TOTAL(1 pulse = fixed mass as set in Fct 4.5.2)
* VOLUME.FLOW(Frequency output 0 to MAX Freq. Hz = Volume flow in range:
MIN. V.FLOW to MAX. V.FLOW as set in Fct. 4.5.2 and Fct. 4.5.3)
* VOL.TOTAL(1 pulse = fixed volume as set in Fct 4.5.2)
* TEMPERAT. (Frequency output 0 to MAX Freq. Hz = Temperature in range:
MIN. TEMP to MAX. TEMP as set in Fct. 4.5.2 and 4.5.3)

CONC. FLOW
CONC. TOTAL
CONC.BY.MASS
CON:BY:VOL.

Concentration parameters if option installed.
See separate instruction manual.

4.5.1 FUNCTION

* DIRECTION (Negative flow gives output of 0 volts DC, Positive flow gives output
of +V volts DC)
* ADDITIONAL(1 pulse = fixed mass as set in Fct 4.5.2)
Value of measured quantity that corresponds to 0 Hz output or

4.5.2

or

LOW LIMIT

PULSE.WIDTH

Minimum Width in mS of Pulse

for functions MASS TOTAL, VOL. TOTAL or CONC.

TOTAL
Value of measured quantity that corresponds to Max. Frequency or

4.5.3

or

HIGH LIMIT

PULSE VAL.

Mass or volume per pulse value

for functions MASS TOTAL, VOL. TOTAL or CONC.

TOTAL

4.5.4 MAX

FREQ

Value of max. frequency quantity that corresponds to the max. measured value
Not accessible for functions OFF, MASS TOTAL, VOL. TOTAL or CONC. TOTAL