KROHNE OPTIMASS with MFC 05x EN User Manual
Page 64
Installation and Operating Instructions OPTIMASS
63
Fct.
No
Text Description
and
settings
4.5
PULSE OUT.
Submenu 4.5 Pulse/frequency output
Function pulse/frequency output P
* OFF (Output = 0V DC)
* MASS FLOW (Frequency output 0 to MAX Freq. Hz = Mass Flow in range:
MIN. FLOW to MAX FLOW as set in Fct. 4.5.2 and 4.5.3)
* DENSITY (Frequency output 0 to MAX Freq. Hz = Density in range:
MIN.DENSITY to MAX.DENSITY as set in Fct. 4.5.2 and 4.5.3)
* MASS TOTAL(1 pulse = fixed mass as set in Fct 4.5.2)
* VOLUME.FLOW(Frequency output 0 to MAX Freq. Hz = Volume flow in range:
MIN. V.FLOW to MAX. V.FLOW as set in Fct. 4.5.2 and Fct. 4.5.3)
* VOL.TOTAL(1 pulse = fixed volume as set in Fct 4.5.2)
* TEMPERAT. (Frequency output 0 to MAX Freq. Hz = Temperature in range:
MIN. TEMP to MAX. TEMP as set in Fct. 4.5.2 and 4.5.3)
CONC. FLOW
CONC. TOTAL
CONC.BY.MASS
CON:BY:VOL.
Concentration parameters if option installed.
See separate instruction manual.
4.5.1 FUNCTION
* DIRECTION (Negative flow gives output of 0 volts DC, Positive flow gives output
of +V volts DC)
* ADDITIONAL(1 pulse = fixed mass as set in Fct 4.5.2)
Value of measured quantity that corresponds to 0 Hz output or
4.5.2
or
LOW LIMIT
PULSE.WIDTH
Minimum Width in mS of Pulse
for functions MASS TOTAL, VOL. TOTAL or CONC.
TOTAL
Value of measured quantity that corresponds to Max. Frequency or
4.5.3
or
HIGH LIMIT
PULSE VAL.
Mass or volume per pulse value
for functions MASS TOTAL, VOL. TOTAL or CONC.
TOTAL
4.5.4 MAX
FREQ
Value of max. frequency quantity that corresponds to the max. measured value
Not accessible for functions OFF, MASS TOTAL, VOL. TOTAL or CONC. TOTAL