Installation and Operating Instructions OPTIMASS

26

1.3.6 Heating and insulation
All secondary containment and heated jacket parts are 316L, except the 1/4" NPT Female connections,

which are 316.

Max heating medium pressure and temperature is 10 bar at 150°C or 145 psig at 300°F.

The max secondary containment pressure on the OPTIMASS 3000 (7100) when fitted with a heating

jacket is 10 bar at 150°C or 145 psig at 300°F.

WITH ¼”NPT(M) INLET/OUTLET

CONNECTIONS

¼” NPF (F) HEATED
JACKET CONNECTION

WITH ¼”NPT(M) INLET/OUTLET

CONNECTIONS

¼” NPT (F) HEATED

JACKET CONNECTION


1.3.7 Purge Port Meters and Burst Disk Meters

Purge Port Options

If the purge port option was selected at time of order, then your meter will be fitted with 1/4" NPT female

connections – these will be clearly identified. These connections are sealed with NPT plugs and PTFE

tape.


Important:

Do not remove these plugs.

The meter is factory sealed with a dry nitrogen gas fill and any ingress of moisture will damage the

meter. The plugs should only be removed to purge the inside of the meter case of any product if it is

suspected that the primary measuring tube has failed. This must only be done after the meter has been

depressurised and removed from service. This should be done as soon as possible after failure is

suspected (less than 3 days)


Burst Disk meters

OPTIMASS 3000 (7100) meters that have been ordered with a bursting (rupture) disk will be so fitted.

This is fitted when the operating pressure of the measuring tube exceeds the design pressure of the

secondary containment. The disk failure pressure is 20bar @ 20°C.



Important:

The burst disk is suitable for the designed application according to the process conditions and flow rates

as per original order. If conditions alter, consult KROHNE for further advice regarding suitability of disk

fitted.

If the product is in any way hazardous, it is strongly recommended that an exhaust tube is connected to

the NPT male thread of the burst disk so that the discharge can be piped to a safe area. This tube

should be large enough that pressure cannot build up in the meter case.

Ensure arrow on burst disk is pointing away from meter.