Installation conditions
Installation conditions
Installation conditions
Installation conditions
Installation
For detailed information refer to
Installation
on page 18.
Inlet run
Minimal 5 DN (straight inlet)
If no details are known, minimal 10 DN recommended
Outlet run
Minimal 3 DN (straight outlet)
If no details are known, minimal 5 DN recommended
Dimensions and weights
For detailed information refer to
Dimensions and weights
on page 105.
Materials
Materials
Materials
Materials
Measuring sensor
Measuring sensor
Measuring sensor
Measuring sensor
Flanges
(wetted)
DN25...65 / 1"...2.5 ": Stainless steel 1.4404 (AISI 316L)
DN80...3000 / 3"...120": Carbon steel
Other materials on request.
Measuring Tube
(wetted)
DN25...300 / 1"...12": Stainless steel 1.4404 (AISI 316L),
some 316Ti / 1.4571
DN350..3000 / 14"...120": Carbon steel
Other materials on request.
Measuring sensor housing
DN25...65 / 1"...2.5": Stainless steel 1.4404 (AISI 316L)
DN80...3000 / 3"...120": Carbon steel
Transducer
Transducers (wetted)
Stainless steel 1.4404 (AISI 316L)
Other materials on request.
Transducer holders
incl. caps
DN350...3000 / 14"...120"; Stainless steel 1.4404 (AISI
316L) (same material as flanges)
Tube transducer cabling
Stainless steel 1.4404 (AISI 316L)
Connection box and connection box support
(remote version only)
Standard: Die-cast aluminium; polyurethane coated
Option: Stainless steel 316 (1.4408)
Coating (measuring sensor)
Standard: Polyurethane
Option: Offshore coating
NACE conformity
On request; wetted materials conform NACE MR 175/103
Signal converter
Signal converter
Signal converter
Signal converter
Housing
Versions C and F: Die-cast aluminium
Option: Stainless steel 316 (1.4408)
Coating
Standard: Polyurethane
Option: Offshore coating