4

ELECTRICAL CONNECTIONS

48

OPTISONIC 3400

www.krohne.com

11/2013 - 4002037702 - HB OPTISONIC 3400 -en-R02

Status output / limit switch passive, modular I/Os

• Any connection polarity.
• U

ext

= 32 VDC

• I ≤ 100 mA
• R

L, max

= 47 kΩ

R

L, min

= (U

ext

- U

0

) / I

max

• open:

I ≤ 0.05 mA at U

ext

= 32 VDC

closed:
U

0, max

= 0.2 V at I ≤ 10 mA

U

0, max

= 2 V at I ≤ 100 mA

• The output is open when the device is de-energized.
• X designates the connection terminals A, B or D, depending on the version of the signal

converter.

Status output / limit switch S

N

NAMUR, modular I/Os

• Any connection polarity.
• Connection in conformity with EN 60947-5-6
• open:

I

nom

= 0.6 mA

closed:
I

nom

= 3.8 mA

• The output is open when the device is de-energized.
• X designates the connection terminals A, B or D, depending on the version of the signal

converter.

Figure 4-18: Status output / limit switch passive S

p

Figure 4-19: Status output / limit switch S

N