11/2013 - 4002037702 - HB OPTISONIC 3400 -en-R02

Status output / limit switch

Status output / limit switch

Status output / limit switch

Status output / limit switch

Function and settings

Adjustable as automatic measuring range conversion,

display of flow direction, overflow, error, switching point

Valve control with activated dosing function

Operating data

Basic I/Os

Modular I/Os

Ex i

Active

-

U

int

= 24 VDC

I ≤ 20 mA

R

L, max

= 47 kΩ

open:

I ≤ 0.05 mA

closed:

U

0, nom

= 24 V at

I = 20 mA

-

Passive

U

ext

≤ 32 VDC

I ≤ 100 mA

R

L, max

= 47 kΩ

R

L, max

= (U

ext

-

U

0

) / I

max

open:

I ≤ 0.05 mA at

U

ext

= 32 VDC

closed:

U

0, max

= 0.2 V at

I ≤ 10 mA

U

0, max

= 2 V at

I ≤ 100 mA

U

ext

= 32 VDC

I ≤ 100 mA

R

L, max

= 47 kΩ

R

L, max

= (U

ext

-

U

0

) / I

max

open:

I ≤ 0.05 mA at

U

ext

= 32 VDC

closed:

U

0, max

= 0.2 V at

I ≤ 10 mA

U

0, max

= 2 V at

I ≤ 100 mA

-

NAMUR

-

Passive to

EN 60947-5-6

open:

I

nom

= 0.6 mA

closed:

I

nom

= 3.8 mA

Passive to

EN 60947-5-6

open:

I

nom

= 0.43 mA

closed:

I

nom

= 4.5 mA

U

i

= 30 V

I

i

= 100 mA

P

i

= 1 W

C

i

= 10 nF

L

i

= 0 mH