background image

6

OPERATION

64

OPTISONIC 3400

www.krohne.com

11/2013 - 4002037702 - HB OPTISONIC 3400 -en-R02

A4 Station Address

A4

Station Address

For Profibus / FF / Modbus devices.

A5 Digital Outputs

A5

Digital Outputs

Valid for all pulse outputs (terminals A, B and/or D) and totaliser 1.

A5.1

Measurement

1) Select measurement: Volume Flow / Mass Flow

2) Use for all outputs? (also use this setting for Fct. A5.2...A5.5!)

Setting: No (only for pulse output D) / Yes (for all digital outputs)

A5.2

Pulse Value Unit

Selection of the unit from a list, depending on the measurement.

A5.3

Value p. Pulse

1) Setting for pulse output D (volume or mass value per pulse)

Setting: xxx.xxx in l/s or kg/s

2) Use for all outputs? Make setting, see Fct. A5.1 above!

A5.4

Low Flow Cutoff

1) Setting for pulse output D (sets output value to "0")

Setting: x.xxx ± x.xxx% (range: 0.0…20%)

(1st value = switching point / 2nd value = hysteresis),

condition: 2nd value ≤ 1st value

2) Use for all outputs? Make setting, see Fct. A5.1 above!

A6 GDC IR interface

A6

GDC IR Interface

After this function has been activated an optical GDC adapter can be

connected to the LC display. After 60 seconds pass without a connection

being established or after the adapter is removed, the function is exited and

the optical keys are active again.

Break (exit function without connection)

Activate (the IR interface adapter and interrupt the optical keys)

No.

Function

Setting / Description