OPERATION

6

89

OPTISONIC 3400

www.krohne.com

11/2013 - 4002037702 - HB OPTISONIC 3400 -en-R02

I

S Electr: Power

Failure

Tot 1 Power Failure

Power failure has occurred.

The totaliser state may be

invalid.

Check the value of the

totaliser.

Tot 2 Power Failure

Tot 3 Power Failure

Power Failure Detected

I

I Electr.

Operation Info.

Zero Calibr. Running

A zero calibration in

progress.

Wait until finished

Sensor Starting up.

Sensor starts up. This is

normal operation at the

beginning of measuring

mode. Other error

messages are suppressed

After some moments, the

converter will act and

respond with the converter

status.

PROFIBUS: no data

No data exchange via

PROFIBUS.

Tot 1 Stopped

Totaliser 1 was stopped.

If totaliser is to continue

counting, select "Yes" in

Fct. C.y.9 (Start Totaliser).

Tot 2 Stopped

Totaliser 2 was stopped.

Tot 3 Stopped

Totaliser 3 was stopped.

Control In A Active

Control In B Active

Status Out A Active

Status Out B Active

Status Out C Active

Status Out D Active

Disp. 1 Overrange

The value in the 1st

measurement line of the

display page is limited.

Check setting for 1st

measurement line.

Disp. 2 Overrange

The value in the 2nd

measurement line of the

display page is limited.

Check setting for 2nd

measurement line.

Optical Interf. Active

The optical interface is being

used. The optical keys are

deactivated.

The keys are ready for

operation again approx. 60

seconds after the end of the

data transfer/removal of the

optical interface.