Description of hart interface, 3 connection variants – KROHNE OPTISONIC 3400 EN User Manual
Page 116
9
DESCRIPTION OF HART INTERFACE
116
OPTISONIC 3400
www.krohne.com
11/2013 - 4002037702 - HB OPTISONIC 3400 -en-R02
9.3 Connection variants
The signal converter is a 4-wire device with 4...20 mA current output and HART
®
interface.
Depending on the version, the settings and the wiring, the current output can operate as passive
or active output.
• Multi-Drop Mode is supported
Multi-Drop Mode is supported
Multi-Drop Mode is supported
Multi-Drop Mode is supported
In a Multi-Drop communication system, more than 2 devices are connected to a common
transmission cable.
• Burst Mode is not supported
Burst Mode is not supported
Burst Mode is not supported
Burst Mode is not supported
In the Burst Mode a slave device transfers cyclic pre-defined response telegrams, to get a
higher rate of data transfer.
There are two ways of using the HART
®
communication:
• as Point-to-Point connection and
• as multi-drop connection, with 2-wire connection or
as multi-drop connection, with 3-wire connection.
INFORMATION!
For detailed information about the electrical connection of the signal converter for HART
®
, refer
to the section "Electrical connection".