Operation – KROHNE OPTISONIC 3400 EN User Manual
Page 87

OPERATION
6
87
OPTISONIC 3400
www.krohne.com
11/2013 - 4002037702 - HB OPTISONIC 3400 -en-R02
S
S Configuration
PROFIBUS Uncertain
IO A Overrange
The output value is limited
by a filter.
Check the range setting of
the output.
IO B Overrange
IO C Overrange
IO D Overrange
S
S Process
Mass Flow Out of Range
The flow is out of range. The
actual flow is higher than
the displayed value.
Check process conditions.
Vol. Flow Out of Range
Velocity Out of Range
M
M Sensor
M
M Electronics
Backplane Data Faulty
The backplane data record is
faulty.
Check the correct
installation of the signal
converter electronics. After
changing one parameter,
the message should
disappear within one
minute. If not, contact the
manufacturer.
Factory Data Faulty
Factory settings are invalid.
Contact manufacturer.
Backplane Difference
The data on the backplane
differ from the data in the
device.
After changing one
parameter, the message
should disappear within one
minute. If not, contact the
manufacturer.
PROFIBUS Baudrate
The PROFIBUS searches for
the current Baudrate.
M
M
Configuration
Backup 1 Data Faulty
Error when checking the
data record for Backup 1.
Use "Setup > Device >
Special Functions > Save
Settings" to save the data
record. If the message
continues to appear, contact
the manufacturer.
Backup 2 Data Faulty
Error when checking the
data record for Backup 2.
M
M Process
F
F
F
F
F Proc: Current
Input