OPERATION

6

87

OPTISONIC 3400

www.krohne.com

11/2013 - 4002037702 - HB OPTISONIC 3400 -en-R02

S

S Configuration

PROFIBUS Uncertain

IO A Overrange

The output value is limited

by a filter.

Check the range setting of

the output.

IO B Overrange

IO C Overrange

IO D Overrange

S

S Process

Mass Flow Out of Range

The flow is out of range. The

actual flow is higher than

the displayed value.

Check process conditions.

Vol. Flow Out of Range

Velocity Out of Range

M

M Sensor

M

M Electronics

Backplane Data Faulty

The backplane data record is

faulty.

Check the correct

installation of the signal

converter electronics. After

changing one parameter,

the message should

disappear within one

minute. If not, contact the

manufacturer.

Factory Data Faulty

Factory settings are invalid.

Contact manufacturer.

Backplane Difference

The data on the backplane

differ from the data in the

device.

After changing one

parameter, the message

should disappear within one

minute. If not, contact the

manufacturer.

PROFIBUS Baudrate

The PROFIBUS searches for

the current Baudrate.

M

M

Configuration

Backup 1 Data Faulty

Error when checking the

data record for Backup 1.

Use "Setup > Device >

Special Functions > Save

Settings" to save the data

record. If the message

continues to appear, contact

the manufacturer.

Backup 2 Data Faulty

Error when checking the

data record for Backup 2.

M

M Process

F

F

F

F

F Proc: Current

Input