ELECTRICAL CONNECTIONS

4

53

OPTISONIC 3400

www.krohne.com

11/2013 - 4002037702 - HB OPTISONIC 3400 -en-R02

Status output/limit switch S

N

NAMUR, Ex i I/Os

• Connection acc. to EN 60947-5-6
• open:

I

nom

= 0.43 mA

closed:
I

nom

= 4.5 mA

• The output is closed when the device is de-energized.
• X designates the connection terminals B or D, depending on the version of the signal

converter.

4.7.6 HART

®

connection

HART

®

connection active (point-to-point)

The parallel resistance to the HART

®

communicator must be R ≥ 230 Ω.

INFORMATION!

Any connection polarity.

Figure 4-27: Status output/limit switch S

N

to NAMUR EN 60947-5-6 Exi

INFORMATION!

In the basic I/O the current output at connection terminals A+/A-/A always has HART

®

capability.

For modular I/O and Ex i E/A, only the output module for the connection terminals C/C- has
HART

®

capability.

Figure 4-28: HART

®

connection active (I

a

)

1 Basic I/O: terminals A and A+
2 Modular I/O: terminals C- and C
3 HART

®

communicator