6

OPERATION

72

OPTISONIC 3400

www.krohne.com

11/2013 - 4002037702 - HB OPTISONIC 3400 -en-R02

C2._ Control Input X

C2._

Control Input X

C2._.1

Mode

X stands for connection terminal A or B

Off (control input switched off) /

Hold All Outputs (hold current values, not display and totalisers) /

Output Y (hold current values) /

All Outputs To Zero (current values = 0%, not display and totalisers) /

Output Y To Zero (current value = 0%) /

All Totalisers (reset all totalisers to "0") /

Totaliser "Z" Reset (set totaliser 1, (2 or 3) to "0") /

Stop All Totalisers /

Stop Totaliser "Z" (stops totaliser 1, (2 or 3) /

Zero Outp.+Stop Tot. (all outputs 0%, stop all totalisers, not the display) /

External Range Y (control input for external range of current output Y) - also

make this setting on current output Y (no check if current output Y is

available) /

Error Reset (all resettable errors are deleted)

Zero Calibration

C2._.2

Invert Signal

Select:

Off (activated output: switch closed) /

On (activated output: switch open)

C2._.3

Information

Serial no. of the I/O board, software version no. and production date of the

circuit board

C2._.4

Simulation

Sequence see B1._ Control Input X

No.

Function

Settings / descriptions

C3 I/O Totalisers

C3.1

Totaliser 1

Set function of totaliser.

_ stands for 1, 2, 3 (= Totaliser 1, 2, 3)

The basic version (standard) has only 2 totaliser!

C3.2

Totaliser 2

C3.3

Totaliser 3

C3._.1

Totaliser Function

Select:

Absolute Total (counts positive + negative values) /

+Totaliser (counts only the positive values) /

-Totaliser (counts only the negative values) /

Off (Totaliser is switched off)

C3._.2

Measurement

Selection of the measurement for Totaliser _

Select:

Volume Flow / Mass Flow

C3._.3

Low Flow Cutoff

Sets the measurement to "0" for low values

Range: 0.0…20%

(1st value = switching point / 2nd value = hysteresis),

condition: 2nd value ≤ 1st value

C3._.4

Time Constant

Range: 000.1…100 s

C3._.5

Preset Value

If this value is reached, positive or negative, a signal is generated that can be

used for a status output at which "Preset Totaliser X" has to be set.

Preset value (max. 8 digits) x.xxxxx in selected unit, see C6.7.10 + 13

C3._.6

Reset Totaliser

Sequence see Fct. A3.2, A3.3 and A3.4