Operation – KROHNE OPTISONIC 3400 EN User Manual
Page 85
OPERATION
6
85
OPTISONIC 3400
www.krohne.com
11/2013 - 4002037702 - HB OPTISONIC 3400 -en-R02
Fieldbus Config.
Check Fieldbus
configuration or load factory
settings.
PROFIBUS Config.
Check PROFIBUS settings or
load factory settings.
Tot 1 FB2 Unit Error
Totaliser is out of operation
due to inadmissible unit.
Check unit in totaliser 1 FB2
or load factory settings.
Tot 2 FB3 Unit Error
Check unit in totaliser 2 FB3
or load factory settings.
Tot 3 FB4 Unit Error
Check unit in totaliser 3 FB4
or load factory settings.
Modbus Config.
Check Modbus configuration
or load factory settings.
Display Config.
Inadmissible settings for the
display
Check display settings or
load factory settings.
IO1 Configuration
Inadmissible settings for IO1 Check settings for IO1 or
load factory settings.
IO2 Configuration
Inadmissible settings for IO2 Check settings for IO2 or
load factory settings.
Tot 1 Configuration
Inadmissible settings for
totaliser 1
Check settings for totaliser 1
or load factory settings.
Tot 2 Configuration
Inadmissible settings for
totaliser 2
Check settings for totaliser 2
or load factory settings.
Tot 3 Configuration
Inadmissible settings for
totaliser 3
Check settings for totaliser 3
or load factory settings.
IO A Configuration
Inadmissible settings for IO
A
Check settings for IO A or
load factory settings.
IO B Configuration
Inadmissible settings for IO
B
Check settings for IO B or
load factory settings.
IO C Configuration
Inadmissible settings for IO
C
Check settings for IO C or
load factory settings.
IO D Configuration
Inadmissible settings for IO
D
Check settings for IO D or
load factory settings.
F
F
F
F
F Process
C
C Sensor
C
C Electronics
Error
type
Event group
Single event
Description
Actions to eliminate the
event