6

OPERATION

70

OPTISONIC 3400

www.krohne.com

11/2013 - 4002037702 - HB OPTISONIC 3400 -en-R02

C2._.10

Invert Signal

Select:

Off (activated output: switch closed) /

On (activated output: switch open)

C2._.11

Phase Shift w.r.t. B

Only available when configuring the A or D terminal and only if output B is a

pulse or frequency output. If setting in Fct. C2.2.7 is "Both Polarities", the

phase shift is prefixed by a symbol, e.g. -90° and +90°

Select:

Off (no phase shift) /

0° Phase Shift (between outputs A or D and B, inversion possible) /

90° Phase Shift (between outputs A or D and B, inversion possible) /

180° Phase Shift (between outputs A or D and B, inversion possible)

C2.3.11

Special Functions

This function is only available at the pulse output of terminal B. At the same

time, 2 pulse outputs must be available:

1st output at terminal A or D / 2nd output at terminal B

The B output is operated as a slave output, controlled and set using master

output A or D

Select:

Off (no phase shift) /

Phase Shift w.r.t. D or A (slave output is B and master output is D or A)

C2._.12

Information

Serial no. of the I/O board, software version no. and production date of the

circuit board

C2._.13

Simulation

Sequence see B1._ Pulse Output X

C2._ Status Output X

C2._

Status Output X

X (Y) stands for one of the connection terminals A, B, C or D

C2._.1

Mode

The output shows the following measuring conditions:

Out Of Specification (output set, signals status of category "Error in Device"

or "Application Failure" or "Out Of refer to

Status messages and diagnostic

information

on page 83 )/

Application Failure (output set, signals status of category "Error in Device"

or "Application Failure" refer to

Status messages and diagnostic

information

on page 83) /

Flow Polarity (polarity of the current flow)

Flow Over Range (over range of the flow)

Totaliser 1 Preset (activates when totaliser X preset value is reached) /

Totaliser 2 Preset (activates when totaliser X preset value is reached) /

Totaliser 3 Preset (activates when totaliser X preset value is reached) /

Output A (activated by the status of output Y, additional output data see

below) /

Output B (activated by the status of output Y, additional output data see

below) /

Output C (activated by the status of output Y, additional output data see

below) /

Output D (activated by the status of output Y, additional output data see

below) /

Off (switched off) /

Empty Pipe (when pipe empty, output activated) /

Error in Device (output set, signals status of category "Error in Device" refer

to

Status messages and diagnostic information

on page 83)

C2._.2

Current Output Y

Only appears if output A...C is set under "Mode" (see above), and this output

is a "Current Output".

Polarity (is signalled)

Over Range (is signalled)

Automatic Range signals lower range

C2._.2

Frequency Output Y and

Pulse Output Y

Only appears if output A, B or D is set under "Mode" (see above), and this

output is a "Frequency/Pulse Output".

Polarity (is signalled)

Over Range (is signalled)

No.

Function

Settings / descriptions