Technical data, 2 technical data – KROHNE OPTISONIC 3400 EN User Manual
Page 93
TECHNICAL DATA
8
93
OPTISONIC 3400
www.krohne.com
11/2013 - 4002037702 - HB OPTISONIC 3400 -en-R02
8.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).
Measuring system
Measuring principle
Ultrasonic transit time
Application range
Flow measurement of (non) conductive fluids
Measured value
Measured value
Measured value
Measured value
Primary measured value
Transit time
Secondary measured values
Volume flow, mass flow, flow speed, flow direction,
velocity of sound, gain, signal to noise ratio, reliability of
flow measurement, totalised volume or mass
Design
Features
3 parallel acoustic path fully welded.
Modular construction
The measurement system consists of a measuring sensor
and a signal converter.
Compact version
OPTISONIC 3400
Remote version
OPTISONIC 3000 F with UFC 400 signal converter
Nominal diameter
DN25...3000 / 1...120"
Measurement range
0.3...20 m/s / 0.98...65 ft/s
Signal converter
Signal converter
Signal converter
Signal converter
Inputs / outputs
Current (incl. HART
®
), pulse, frequency and/or status
output, limit switch and/or control input (depending on
the I/O version)
Totaliser
2 (optional 3) internal totalisers with a max. of 8 digits
(e.g. for totalising volume and/or mass units)
Verification and self-diagnostics
Integrated verification, diagnostic functions: measuring
device, process, measured values, device configuration,
etc.
Communication interfaces
Modbus RS485, HART
®
7, Foundation Fieldbus ITK6,
Profibus PA Profile 3.02