ELECTRICAL CONNECTIONS

44

OPTISONIC 3400

www.krohne.com

11/2013 - 4002037702 - HB OPTISONIC 3400 -en-R02

4.7.4 Modular inputs/outputs and bus systems

Current output active (only current output terminals C/C- have HART

®

capability),

modular I/Os

• U

int, nom

= 24 VDC

• I ≤ 22 mA
• R

L

≤ 1 kΩ

• X designates the connection terminals A, B or C, depending on the version of the signal

converter.

Current output passive (only current output terminals C/C- have HART

®

capability),

modular I/Os

• U

ext

≤ 32 VDC

• I ≤ 22 mA
• U

0

≥ 1.8 V

• R

L, max

= (U

ext

- U

0

) / I

max

• X designates the connection terminals A, B or C, depending on the version of the signal

converter.

CAUTION!
Observe connection polarity.

INFORMATION!

For further information on electrical connection refer to Description of the inputs and outputs
on page 35

.

For the electrical connection of bus systems, please refer to the separate documentation for
the respective bus systems.

INFORMATION!

For frequencies above 100 Hz, shielded cables are to be used in order to reduce effects from
electrical interferences (EMC).

Compact and field housing versions:

Compact and field housing versions:

Compact and field housing versions:

Compact and field housing versions: Shield connected via the cable terminals in the terminal
compartment.

Any connection polarity.

Figure 4-12: Current output active I

a