ELECTRICAL CONNECTIONS

50

OPTISONIC 3400

www.krohne.com

11/2013 - 4002037702 - HB OPTISONIC 3400 -en-R02

Control input active C

N

NAMUR, modular I/Os

• Connection acc. to EN 60947-5-6
• Switching point for identifying "contact open or closed":

Contact open (off): U

0, nom

= 6.3 V with I

nom

< 1.9 mA

Contact closed (on): U

0, nom

= 6.3 V with I

nom

> 1.9 mA

• Detection of cable break:

U

0

≥ 8.1 V with I ≤ 0.1 mA

• Detection of cable short circuit:

U

0

≤ 1.2 V with I ≥ 6.7 mA

• X designates the connection terminals A or B, depending on the version of the signal

converter.

4.7.5 Ex i inputs/outputs

CAUTION!
Observe connection polarity.

Figure 4-22: Control input active C

N

to NAMUR EN 60947-5-6

DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.

INFORMATION!
For further information on electrical connection refer to Description of the inputs and outputs on
page 35

.

INFORMATION!

For frequencies above 100 Hz, shielded cables are to be used in order to reduce effects from
electrical interferences (EMC).

Compact and field housing versions:

Compact and field housing versions:

Compact and field housing versions:

Compact and field housing versions: Shield connected via the cable terminals in the terminal
compartment.

Any connection polarity.