Electrical connections, 2 current output – KROHNE OPTISONIC 3400 EN User Manual

Page 36

background image

4

ELECTRICAL CONNECTIONS

36

OPTISONIC 3400

www.krohne.com

11/2013 - 4002037702 - HB OPTISONIC 3400 -en-R02

4.6.2 Current output

• All outputs are electrically isolated from each other and from all other circuits.
• All operating data and functions can be adjusted.
• Passive mode: external power U

ext

≤ 32 VDC at I ≤ 22 mA

• Active mode: load impedance R

L

≤ 1 kΩ at I ≤ 22 mA;

R

L

≤ 450 Ω at I ≤ 22 mA for Ex i outputs

• Self-monitoring: interruption or load impedance too high in the current output loop
• Error message possible via status output, error indication on LC display.
• Current value error detection can be adjusted.
• Automatic range conversion via threshold or control input. The setting range for the

threshold is between 5 and 80% of Q

100%

, ± 0...5% hysteresis (corresponding ratio from

smaller to larger range of 1:20 to 1:1.25).
Signaling of the active range possible via a status output (adjustable).

• Forward / reverse flow measurement (F/R mode) is possible.

INFORMATION!
The current outputs must be connected depending on the version! Which I/O versions and
inputs/outputs are installed in your signal converter are indicated on the sticker in the cover of
the terminal compartment.

INFORMATION!
For further information refer to Connection diagrams of inputs and outputs on page 39

.

DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.