Jenway 6715 User Manual
Page 96

78
Selecting Sample Style
This option allows the user to select to measure one sample at up to four wavelengths, or to measure
up to four samples at different wavelengths when the 8-position cell holder is fitted. This parameter is
used in conjunction with setting wavelengths; i.e. in multiple option Sample 1 = Wavelength 1,
Sample 2 = Wavelength 2, etc, with position 0 reserved for a common blank.
Single – allows a single sample to be measured at up to 4 wavelengths.
Multiple – allows up to 4 samples to be measured at single different wavelengths.
Method Security
The 67 Series spectrophotometers can support up to 10 individual users plus one Supervisor who has
full administrative rights.
Logged in users can create methods with three levels of security options:
Personal – these methods are only accessible by the originator.
Public – these methods are available for use and modification by any logged in user.
Read-Only – these methods can be accessed by all logged in users, but can only be modified by the
originator.
The preferred level of protection can be achieved by selecting the
Method Security key that toggles
between
Personal, Public and Read-Only options.