Jenway 6715 User Manual
Page 101

83
If the current user does not have the required privileges to delete the selected method then an
information message will be displayed advising that the method cannot be deleted.
Privileges required for deleting designated methods:
Public Methods – only the Supervisor and Originator can delete these.
read-only Methods – only the Supervisor and Originator can delete these.
Personal Methods – Only the Originator can delete these methods. (The Supervisor can delete these
by re-setting the Originator’s PIN code and then logging in as the Originator).
Saving results
After completion of a measurement the result can be saved by simply pressing the
Save key on the
display surround. The result is saved under the method that created it, with the entered Batch ID and
an incremental number along with the time and date of the measurement. Results can also be saved
as part of the
Auto Log function by selecting Memory under Auto Log options, which will vary
depending on the type of sampling accessory fitted.
Printing results
After completion of a measurement the result can be printed, by simply pressing the
Print key on the
display surround. The result will be printed to either the internal or external printer, as selected by the
user in the
Printer Settings option.
The first result of any new batch is preceded by a print header, which gives details of the method
settings and Batch ID. Results can also be printed as part of the
Auto Log function, which will vary
depending on the type of sampling accessory fitted.
recalling results
Stored results are always directly linked to the method that created them. To access results first recall
the method as described in
Recalling Methods. With the method highlighted touch the Search
Results icon. This will open a screen detailing all results available to the current user. Touch the
required result or batch (depending on the mode) and then the
Open Specific Result icon. This
will display the results on the screen. The
Tools option can then be used to work on these results
(depending on mode). It is also possible to print the result by simply pressing the
Print key on the
display surround. Options to print to the Internal or External printer or to the CSV file will be displayed.
Printing to the CSV (Comma Separated Values) file will save the data in CSV format on the external
data card. (If a card is not fitted the instrument will display a prompt). This is in text format and can be
viewed and printed in Excel
®
.
In
Kinetics mode an additional option is available (Analogue Mode). This allows results to be output
via the rear panel analogue sockets to a chart recorder or similar device.
Sharing results
Results attached to
Personal Methods cannot be accessed by any other user.
Results attached to
Read-Only and Public Methods can be accessed by all users, based on their
current