Jenway 6715 User Manual
Page 135

117
replicates Setup…
Selecting
Replicates Setup… opens a dedicated sub-menu.
Up to 5 replicates may be selected with the recorded result being the average of the 5 or as the last
measured result.
Setting Method Security
The 67 Series spectrophotometers can support up to 10 individual users plus one Supervisor who has
full administrative rights.
Logged in users can create methods with three levels of security options:
Personal – these methods are only accessible by the originator.
Public – these methods are available for use and modification by any logged in user.
Read-Only – these methods can be accessed by all logged in users, but can only be modified by the
originator.
The preferred level of protection can be achieved by selecting the
Method Security key that toggles
between
Personal, Public and Read-Only options.