Jenway 6715 User Manual
Page 158

140
10.5
Peltier MoDUle
This accessory module is fitted into the instrument casing by following the instructions for the Cell
Accessories.
overview of operation
This temperature controlled sample module uses the electronic peltier heat-transfer effect to maintain
a constant temperature around the sample cuvette, heating or cooling as necessary to maintain
control. A desired sample temperature between 20 and 50°C (or 68 and 122°F) that can be set to
a resolution of 0.1°C (1° F) is available within the Accessory settings on the spectrophotometer.
Additional icons and data will appear in the status bar at all times showing the actual temperature and
the set (target) level. Control is fully automatic and once set is stored within the method, needing no
further entry or adjustment when that method is re-used at a later date.
note: Depending on the insulating/conducting properties of the material used in the
chosen cuvette a varying period of equilibration may be required to bring the actual sample
temperature to that shown on the display.
Warning: Do not cover or obstruct the exhaust air vent on the front of the accessory module,
this will adversely affect the temperature regulation and may damage the peltier element or
control components.
installation
If your instrument was ordered with the Peltier it will be supplied with the module installed.
Fitting the Peltier Module (as a separate accessory option)
If you are changing from an existing sample chamber module to the Peltier the following installation
instructions should be followed:
Disconnect the instrument from the mains supply.
Check the installed sample chamber module and remove any filled cuvettes to avoid spillage.
Raise lid by 30°
for re-assembly
Ensure lugs locate
into main case
Sample chamber
retaining screw