PerForMinG MeASUreMentS – all users
Performing a Scan
Once the required settings have been confirmed the main
Spectrum measurement screen will be
shown with the
Scan key inactive (blanked out):
Performing a Baseline
Insert a cuvette containing a blank solution into the sample holder and close the sample chamber lid.
(Test tubes or other sample containers may be used depending on the sample holder accessory fitted).
Select the
Baseline key. The messages ‘Performing baseline scan. Please wait…’ are displayed
momentarily. Once the baseline measurement and storage is completed the screen will update to
show:
note: In general the blank solution should contain everything that is in the sample except the
colour-producing component. For specific information reference should be made to the procedure or
application being followed. For enhanced reproducibility matched cuvettes should be used.
Scanning the Sample
Once the baseline has been performed the
Scan key will be active. Remove the cuvette containing the
blank solution and insert the sample into the sample holder and close the sample chamber lid. Select
the
Scan key, if Auto Scale has been selected once the scan is completed, the screen will blank and
the scan will be re-drawn to the optimum scaling.