OPERATION

93

4. To stop the test when HOLD is

displayed on-screen, press the
STOP button.

STOP

Note

When in HOLD status, only the START and STOP
buttons can be pressed, all other keys are disabled.

Stop a Running
Test

1. To stop the AUTO test at any time

when it is running, press the STOP
button. The AUTO test will stop
immediately. When the STOP
button is pressed, a judgment is
not made on the current test and
any remaining tests are aborted.

All panel keys except the STOP
and START buttons are locked
when the tester has been stopped.
All the results up until when the
AUTO test was stopped are shown
on-screen. See page 94 for more
details on automatic test results.

STOP

E

M

A U O

T

N A

_

*

A U T O = 0 0 1 - * *

# 0 1 : F A I

# 0 2 : P A S

# 0 3 : S T O

# 0 4 : - - -

# 0 5 : - - -

# 0 6 : - - -

# 0 7 : - - -

# 0 8 : - - -

# 0 9 :

# 1 0 :

# 1 1 :

# 1 2 :

# 1 3 :

# 1 4 :

# 1 5 :

# 1 6 :

L

-

S

-

P

-

-
-

Example of an automatic test that has been
stopped. Dashes (-) indicate aborted test steps.

2. To put the tester back into READY

status, press the STOP button
again.

STOP