Setting pass hold, Operation – GW Instek GPT-9900 series User Manual User Manual

Page 55

background image

OPERATION

55

Setting PASS HOLD

Background

The PASS HOLD settings only apply to the
selected test in an AUTO test. When the PASS
HOLD setting is set to ON, a PASS judgment is
held until the START button is pressed.

Note

The PASS HOLD setting only applies to AUTO
tests. This setting is ignored when running a
single
MANU test.

Steps

1. Press the UTILITY key on the front

panel when the tester is in EDIT
status. The display will go from the
normal EDIT status to the MANU
Utility menu for the current test.

UTILITY

L I T Y

I

M A U

N

U T

2

M A N U = * * * - 0 0

A R C M D E : O

O

F

F

G R O U N D M O D E :

N

O

P A S S H

L D :

O

O F

F

F A I L M D E :

O

P

S T O

M A X H O L D : O F F

Note

The MANU UTILITY settings only apply to the
selected MANU test.

2. Use the UP/DOWN arrow keys to

move to the PASS HOLD setting.

3. Use the scroll wheel to set PASS

HOLD.

PASS HOLD

OFF, ON

4. Press the EDIT/SAVE key to save

and exit the MANU Utility menu.

EDIT/SAVE