REMOTE CONTROL

141

MANU:IR:REF

Set

Query

Description

Sets or returns the IR reference value in MΩ (GPT-
98XX) or GΩ. The test must first be in IR mode
before this command can be used.

The reference value must be lower than the HI SET
value.

Syntax
Query Syntax

MANU:IR:REF
MANU:IR:REF?

Parameter/
Return parameter

GPT-98XX only:
0000 ~ 9999 (unit = MΩ)
GPT-99XX/GPT-99XXA only:
Format A: 0 ~ 50.00 (unit = GΩ)
Format B: 0G ~ 50.00G
Format C: 0M ~ 50000M

Example
(GPT-98XX)

MANU:IR:REF 900
Sets the IR reference to 900 MΩ.

Example
(GPT-99XX/
99XXA)

MANU:IR:REF 0.900
Sets the IR reference to 900 MΩ.

MANU:GB:CURRent

Set

Query

Description

Sets or returns the GB current in A. The test must
first be in GB mode before this command can be
used.

Syntax
Query Syntax

MANU:GB:CURRent
MANU:GB:CURRent?

Parameter/
Return parameter

3.00~33.00 (GPT-98XX)
3.00~33.00 (GPT-99XX)

Example

MANU:GB:CURR 3.00
Sets the GB current to 3.00A.