beautypg.com

System:buzzer:ftime, System:error – GW Instek GPT-9900 series User Manual User Manual

Page 125

background image

REMOTE CONTROL

125

SYSTem:BUZZer:FTIMe

Set

Query

Description

Sets the FAIL Sound duration in seconds.

Syntax
Query Syntax

SYSTem:BUZZer:FTIMe
SYSTem:BUZZer:FTIMe?

Parameter/
Return parameter

0.2~999.9

Example

SYST:BUZZ:FTIM 1
Sets the buzzer to 1 second for a FAIL judgment.

SYSTem:ERRor

Query

Description

Returns any errors in the output buffer. See the
error code table below for details.

Query Syntax

SYSTem:ERRor ?

Return parameter

Returns an error string that includes
an error code and an error
description.

Error Code Table

Error code, Error description

0,No Error

20,Command Error

21,Volume Error

22,String Error

23,Query Error

24,Mode Error

25,Time Error

26,DC Over 50W (GPT-98XX), DC Over 100W (GPT-
99XX/99XXA

27,GBV > 5.4V

30,Voltage Setting Error

31,Current Setting Error

32,Current HI SET Error

33,Current LOW SET Error

34,Resistance HI SET Error