beautypg.com

Manu:acw:chiset, Manu:acw:closet – GW Instek GPT-9900 series User Manual User Manual

Page 132

background image

GPT-9000/9000A Series User Manual

132

MANU:ACW:CHISet

Set

Query

Description

Sets or returns the ACW HI SET current value in
milliamps. The test must first be in ACW mode
before this command can be used.

Syntax
Query Syntax

MANU:ACW:CHISet
MANU:ACW:CHISet?

Parameter/
Return parameter

0.001 ~ 042.0 (GPT-98XX)
0.001 ~ 110.0 (GPT-99XX/99XXA)

Example

MANU:ACW:CHIS 10.0
Sets the ACW HI SET current to 10 mA.

MANU:ACW:CLOSet

Set

Query

Description

Sets or returns the ACW LO SET current value in
milliamps. The LO SET value must be less than the
HI SET value. The test must first be in ACW mode
before this command can be used.

The LO SET range must use the HI SET range. If all
the digits in the LO SET range are outside the HI
SET range, an error will be produced. All digits
outside the HI SET range are ignored and will not
be used.

For example:
HI SET value: 12.34
LO SET value1: 0.005  error
LO SET value2: 0.053  no error
In the example above LO SET value1 will produce
an error as all digits are outside the range of HI
SET. LO SET value2 will not produce an error, but
will return 0.05, not 0.053.