APPENDIX

167

Current Best Resolution

GPT-98XX:
1uA
0.001mA(0.001mA~0.999mA)
0.01mA(01.00mA~09.99mA)
0.1mA(010.0~040.0mA)

GPT-99XX/GPT-99XXA:
1uA
0.001mA(0.001mA~1.100mA)
0.01mA(01.11mA~11.00mA)
0.1mA(011.1~100.0mA)

Current Measurement
Accuracy

GPT-98XX:
± (1.5% of rdg + 30 counts) when HI SET<1.00mA
± (1.5% of rdg + 3 counts) when HI SET≥1.00mA

GPT-99XX/GPT-99XXA:
± (1.5% of rdg + 30 counts) when HI SET<1.11mA
± (1.5% of rdg + 3 counts) when HI SET≥1.11mA

Window Comparator Method Yes
ARC DETECT

Yes

Rise-time Control Function

Yes

RAMP (Ramp Time)

0.1~999.9S

TIMER (Test Time)

OFF*, 0.5S~999.9S

GND

ON/OFF

* The timer can only be turned off under special MANU mode (MANU=***-000)

DC Withstanding Voltage
Output Voltage Range

0.100kV~ 6.000kV

Output Voltage Resolution

2V

Output Voltage Accuracy

(1% of setting +5V) with no load

Maximum Rated Load(Table1) 50W (5kV/10mA)[GPT-98XX]

100W (5kV/20mA)[GPT-99XX/99XXA]

Maximum Rated Current

10mA [GPT-98XX, 20mA [GPT-99XX/99XXA]
0.001mA ~ 2mA (0.1kV≤V≤0.5kV)
0.001mA ~ 10mA (0.5kV0.001mA ~ 20mA (0.5kV99XXA]

Voltmeter Accuracy

(1% of reading+ 5V)

Voltage Regulation

1% +5V

[Maximum rated load → no load]

Current Measurement Range 0.001mA~010.0mA [GPT-98XX]

0.001mA~020.0mA [GPT-99XX/99XXA]

Continued…