beautypg.com

Operation, Id e, Cursor – GW Instek GPT-9900 series User Manual User Manual

Page 45

background image

OPERATION

45

3. Press the HI/LO soft-key again or

press the DOWN arrow key to
switch between HI SET and LO
SET.

I / L

H

O

OR

I R

G B

A CW

D CW

I / L

H

O

I M E

T

R

E

F R E Q =

0 H z

6

0

100

k

V

I

D

E

M

T I

E

= 0 0 1 . 0 S

R

0 0 m A

E

R F

=

#

0 .

0

M

M A U

N

N A

_

1 . 0 0 m A

0

L

E

S

= 0

T

0 0 . 1 S

0

A M P

=

R

T

2

M A N U = * * * - 0 0

cursor

1 . 0 0 m A

I

H

E

S

= 0

T

m A

4. Use the scroll wheel to set the HI

SET/LO SET limit.

ACW (LO)


DCW (LO)


IR (HI)


GB (LO)

0.000mA~041.9mA (GPT-98XX)
0.000mA~109.9mA (GPT-99XX/
99XXA)
0.000mA~010.9mA (GPT-98XX)
0.000mA~020.9mA (GPT-99XX/
99XXA)
0001MΩ~9999MΩ, ∞ (GPT-98XX)
0.001GΩ~50.00GΩ, ∞ (GPT-99XX/
99XXA)
000.0mΩ ~ 649.9mΩ

Note

The LO SET setting is limited by the HI SET
setting. The LO SET limit cannot be greater than
the HI SET limit.

When setting the current, be aware that a
maximum of 200VA can be set for ACW and 50W
for DCW (GPT-98XX) or 500VA and 100W,
respectively for GPT-99XX/99XXA.