REMOTE CONTROL

139

Example

MANU:DCW:ARCC 10
Sets the DCW ARC value to 10mA.

MANU:IR:VOLTage

Set

Query

Description

Sets or returns the IR voltage in kV. The test must
first be in IR mode before this command can be
used.

Syntax
Query Syntax

MANU:IR:VOLTage
MANU:IR:VOLTage?

Parameter/
Return parameter

0.05 ~ 1 (0.05kV to 1kV: steps of .05)
*GPT-99XX/99XXA also includes a
0.125kV point.

Example

MANU:IR:VOLT 1
Sets the IR voltage to 1 kV.

MANU:IR:RHISet

Set

Query

Description

Sets or returns the IR HI SET resistance value in
MΩ (GPT-98XX) or GΩ. The test must first be in IR
mode before this command can be used.

Syntax
Query Syntax

MANU:IR:RHISet |NULL
MANU:IR:RHISet?

Parameter/
Return parameter

GPT-98XX only:
2 ~ 9999 (unit = MΩ)
GPT-99XX/GPT-99XXA only:
Format A: 0.002 ~ 50.00 (unit = GΩ)
Format B: 0.002G ~ 50.00G
Format C: 2M ~ 50000M

NULL

Sets the HI SET value to ∞.

Example
(GPT-98XX)

MANU:IR:RHIS 10
Sets the IR HI SET resistance to 10 MΩ.