OPERATION

57

2. Use the UP/DOWN arrow keys to

move to the FAIL MODE setting.

3. Use the scroll wheel to set FAIL

MODE.

FAIL MODE

CONTINUE, HOLD, STOP

4. Press the EDIT/SAVE key to save

and exit the MANU Utility menu.

EDIT/SAVE

Note

The ESC key can be pressed at any time in the
MANU Utility menu to cancel and exit.

Setting MAX HOLD

Background

The MAX HOLD setting will hold the
maximum current measured in the ACW and
DCW tests or the maximum resistance
measured in IR and GB tests.

Steps

1. Press the UTILITY key on the front

panel when the tester is in EDIT
status. The display will go from the
normal EDIT status to the MANU
Utility menu for the current test.

UTILITY