Re z, Omeasured resistance, Ei v – GW Instek GPT-9900 series User Manual User Manual

Page 74

background image

GPT-9000/9000A Series User Manual

74

4. The ZERO function can be

activated by pressing the
corresponding soft-key in the
READY status. The ZERO soft-key
will be highlighted.

E R O

Z

5. Press the START button to perform

the zero check. The tester will go
into the ZERO status.

START

0

3 01

A

Z E R 0

I R

G B

A CW

D CW

E

F R E Q =

0 H z

6

R

E

Z

G B V = 0 . 6 2 2 V

M

T I

E

= 0 0 3 . 2 S

R

. 0 m Ω

E

R F

=

#

0 0

0

0 0 . 0 m Ω

I

H

E

S

= 3

T

M

M A U

N

N A

_

2

M A N U = * * * - 0 0

O

Measured Resistance

200 7

m Ω

ZERO status

6. When the zero check has finished, the tester

will return back to the VIEW status. The
resistance of the test leads will be automatically
set as the Reference value.

0

3 01

A

I R

G B

A CW

D CW

E

F R E Q =

0 H z

6

E

I

V

G B V = V

M

T I

E

= 0 0 3 . 2 S

R

. 7 m Ω

E

R F

=

#

0 0

2

0 0 . 0 m Ω

I

H

E

S

= 3

T

M

M A U

N

N A

_

2

M A N U = * * * - 0 0

W

m Ω

Reference value

Note

Remember to replace the test leads to the proper
position on the DUT before testing.