Running the Test

1. In special test mode (000), tests are

started and stopped in the same
way as for the normal manual test
mode. See page 64 for details.

Page 64

2. If required, the scroll wheel can be

used to set the voltage level in real-
time as the test is running (this
does not apply to IR or GB tests).

ACW
DCW

0.100kV ~ 5kV
0.100kV ~ 6kV

Results

Test judgments are the same as
those for the normal manual tests.
Please see the PASS/FAIL MANU
Test section for details.

Page 68

View Sweep
Graph

Unlike normal manual tests, the special test
mode also has an option to view the resultant
test as a sweep graph.

This option is only available for the
GPT-9901A/9902A/9903/9903A/9904.

Steps

1. When the test has finished, press

the SWEEP key to view the results
of the sweep in a graph.

W E E

S

P

Graph Test Items:

TEST

Primary

Secondary

ACW

Test voltage

test current

DCW

Test voltage

test current

IR

Test voltage

test resistance

GB

Test current

test resistance