REMOTE CONTROL

135

MANU:ACW:ARCCurrent

Set

Query

Description

Sets or returns the ACW ARC current value in mA.
ARC must be enabled before the ARC current can
be set. The test must first be in ACW mode before
this command can be used.

ARC current uses the same range as the HI SET
value. The ARC current is limited to 2X the HI SET
value.

Syntax
Query Syntax

MANU:ACW:ARCCurrent
MANU:ACW:ARCCurrent?

Parameter/
Return parameter

1.000 ~ 080.0 (GPT-98XX)
2.000 ~ 200.0 (GPT-99XX/99XXA)

Example

MANU:ACW:ARCC 0.04
Sets the ACW ARC value to 0.04 mA.

MANU:DCW:VOLTage

Set

Query

Description

Sets or returns the DCW voltage in kV. The test
must first be in DCW mode before this command
can be used.
Note: A “DC Over 50W” error will result if the
DCW Voltage X HI SET value is > 50 watts (GPT-
98XX).
Note: A “DC Over 100W” error will result if the
DCW Voltage X HI SET value is > 100 watts (GPT-
99XX/99XXA).

Syntax
Query Syntax

MANU:DCW:VOLTage
MANU:DCW:VOLTage?

Parameter/
Return parameter

0.100 ~ 6.100 (kV)

Example

MANU:DCW:VOLT 6
Sets the DCW voltage to 6 kV.