GPT-9000/9000A Series User Manual

134

MANU:ACW:FREQuency

Set

Query

Description

Sets or returns the ACW test frequency in Hz. The
test must first be in ACW mode before this
command can be used.

Syntax
Query Syntax

MANU:ACW:FREQuency {50|60}
MANU:ACW:FREQuency?

Parameter/
Return parameter

50

50 Hz

60

60 Hz

Example

MANU:ACW:FREQ 50
Sets the ACW test frequency to 50Hz.

MANU:ACW:REF

Set

Query

Description

Sets or returns the ACW reference value in mA.
The test must first be in ACW mode before this
command can be used.

The ACW reference value must be less than the HI
SET value.

The ACW reference value must use the same range
as the HI SET value.

Syntax
Query Syntax

MANU:ACW:REF
MANU:ACW:REF?

Parameter/
Return parameter

0.000 ~ 041.9 (GPT-98XX)
0.000 ~ 109.9 (GPT-99XX/99XXA)

Example

MANU:ACW:REF 0.01
Sets the ACW reference to 0.01 mA.