REMOTE CONTROL

137

an error as all digits are outside the range of HI
SET. LO SET value2 will not produce an error, but
will return 0.05, not 0.053.

Syntax
Query Syntax

MANU:DCW:CLOSet
MANU:DCW:CLOSet?

Parameter/
Return parameter

0.000 ~ 010.9 (GPT-98XX)
0.000 ~ 020.9 (GPT-99XX/99XXA)

Example

MANU:DCW:CLOS 2.00
Sets the DCW LO SET current to 2mA.

MANU:DCW:TTIMe

Set

Query

Description

Sets or returns the DCW test time in seconds. The
test must first be in DCW mode before this
command can be used.

In special MANU mode, the TIMER can be turned
off.

Syntax
Query Syntax

MANU:DCW:TTIMe {|OFF}
MANU:DCW:TTIMe?

Parameter


OFF

0.5 ~ 999.9 seconds
TIMER OFF (special MANU mode).

Return parameter

TIME OFF

0.5 ~ 999.9 seconds
TIMER is OFF (special MANU mode).