2. Use the scroll wheel to set the

TIMER value.

ACW
DCW
IR
GB

000.5s~999.9s
000.5s~999.9s
001.0s~999.9s
000.5s~999.9s

Note

With the ACW test function, when the test current
is between 30mA and 40mA (GPT-98XX) or 80mA
and 100mA (GPT-99XX/99XXA), the ramp time +
test time cannot exceed 240 seconds. At this
current level, the tester also needs to pause after a
test for a time equal to or greater than the output
time. See the specifications on page 166 for
details.

Special Manual
Mode

When in special MANU test mode (page 73) the
Timer can be turned off when using the DCW
or ACW test function.

Hold the TIMER soft-key for 3
seconds to turn the timer off.