GPT-9000/9000A Series User Manual

140

Example
(GPT-99XX/
99XXA)

MANU:IR:RHIS 0.010
Sets the IR HI SET resistance to 10 MΩ.

MANU:IR:RLOSet

Set

Query

Description

Sets or returns the IR LO SET resistance value in
MΩ (GPT-98XX) or GΩ. The LO SET value must be
less than the HI SET value. The test must first be
in IR mode before this command can be used.

Syntax
Query Syntax

MANU:IR:RLOSet
MANU:IR:RLOSet?

Parameter/
Return parameter

GPT-98XX only:
1 ~ 9999 (unit = MΩ)
GPT-99XX/GPT-99XXA only:
Format A: 0.001 ~ 50.00 (unit = GΩ)
Format B: 0.001G ~ 50.00G
Format C: 1M ~ 50000M

Example
(GPT-98XX)

MANU:IR:RLOS 10
Sets the IR LO SET resistance to 10MΩ.

Example
(GPT-99XX/
99XXA)

MANU:IR:RLOS 0.010
Sets the IR LO SET resistance to 10MΩ.

MANU:IR:TTIMe

Set

Query

Description

Sets or returns the IR test time in seconds. The test
must first be in IR mode before this command can
be used.

Syntax
Query Syntax

MANU:IR:TTIMe
MANU:IR:TTIMe?

Parameter/
Return parameter

1.0 ~ 999.9 seconds

Example

MANU:IR:TTIM 1
Sets the IR test time to 1 second.