Running a MANU Test

Background

A test can be run when the tester is in READY
status.

Note

The tester cannot start to run a test under the
following conditions:

A protection setting has been tripped; when a
protection setting has been tripped the
corresponding error message is displayed on
the screen. See page 164 for a comprehensive
list of the all the setting errors.

The INTERLOCK function is ON and the
Interlock key is not inserted in the signal I/O
port (page 102).

The STOP signal has been received remotely.

If Double Action is ON, ensure the START
button is pressed immediately after the STOP
button (<0.5s).

Note

When a test is running the voltage output
cannot be changed, unless the test is under the
special manual mode. See page 73 for details.

Steps

1. Ensure the tester is in VIEW status

for the current test. Save the
current test if necessary.