REMOTE CONTROL

133

Syntax
Query Syntax

MANU:ACW:CLOSet
MANU:ACW:CLOSet?

Parameter/
Return parameter

0.000 ~ 041.9 (GPT-98XX)
0.000 ~ 109.9 (GPT-99XX/99XXA)

Example

MANU:ACW:CLOS 20.0
Sets the ACW LO SET current to 20 mA.

MANU:ACW:TTIMe

Set

Query

Description

Sets or returns the ACW test time in seconds. The
test must first be in ACW mode before this
command can be used.

Note: A “TIME ERR” will result if the Ramp Time
+ Test Time is ≥ 240 seconds when the HI SET limit
is over 30mA (GPT-98XX) or over 80mA (GPT-
99XX/99XXA). This applies to the ACW function
only.

In special MANU mode, the TIMER can be turned
off.

Syntax
Query Syntax

MANU:ACW:TTIMe {|OFF}
MANU:ACW:TTIMe?

Parameter


OFF

0.5 ~ 999.9 seconds
TIMER OFF (special MANU mode).

Return parameter

TIME OFF

0.5 ~ 999.9 seconds
TIMER is OFF (special MANU
mode).

Example

MANU:ACW:TTIM 1
Sets the ACW test time to 1 second.