Setting the grounding mode – GW Instek GPT-9900 series User Manual User Manual

Page 58

background image

GPT-9000/9000A Series User Manual

58

2. Use the UP/DOWN arrow keys to

move to the MAX HOLD setting.

3. Use the scroll wheel to set MAX

HOLD.

MAX HOLD

OFF, ON

4. Press the EDIT/SAVE key to save

and exit the MANU Utility menu.

EDIT/SAVE

Note

The ESC key can be pressed at any time in the
MANU Utility menu to cancel and exit.

Setting the Grounding Mode

Background

When GROUND MODE is set to ON, the GPT-
9000/9000A grounds the return terminal to the
ground. This mode is best for DUTs that are
grounded to an earth ground by their chassis,
fixtures or operation environment. This mode
measures the potential of the HIGH VOLTAGE
terminal with respect to earth ground. This
means that any stray capacitance/resistance
that leaks to earth ground will also be
measured. This is the safest testing mode,
though potentially not as accurate.

When GROUND MODE is set to OFF, the
return terminal is floating with respect to the
earth ground. This mode is for DUTs that are
floating and not directly connected to an earth
ground. This is more accurate than when
GROUND MODE is set to ON as any stray
capacitance/resistance that leaks to the earth