Fuse Replacement

Set the SAS/SATA InFusion power switch to OFF and disconnect the power cord before
inspecting or replacing the fuse. Open the black fuse holder using a small, flat-bladed
screwdriver. Remove the old fuse, replace it with a new 5x20 mm "T" rated 1.0 A/250 V
fuse, and reinstall the fuse holder.

Cleaning

Clean only the exterior of the SAS/SATA InFusion, using a damp, soft cloth. Do not use
chemicals or abrasive elements. Under no circumstances allow moisture to penetrate the
instrument. To avoid electrical shock, unplug the power cord from the AC outlet before
cleaning.

WARNING

For continued fire protection at all line voltages, replace fuse
with the specified type and rating only. Disconnect the power
cord before replacing fuse.

WARNING

No operator serviceable parts inside. Do not remove covers.
Refer servicing to qualified personnel.